UJI VALIDITAS SUHU DENGAN PENGATURAN WAKTU PADA AUTOCLAVE STEAM STERILIZER DI RSUD NGANJIK
Abstract
In the medical area, keeping the medical equipment's hygiene is the most important
thing to do. All of the equipment related to the patient including to do an operation, before
they're used, it" must be cleaned first. If not, it cans contagious disease to another patient, the
hospital's visitors, even the medicalojjicer itself. In the point, the function of the hospital is a
place to heal the sick people. But hospital can be also become the source of disease because
there are many patient bringing much disease in the hospital, both catching disease and not
catching disease. In spite of that, the sterilization process in a hospital is needed and a must
which have 10 do like the fixed procedure that have been specified.
Related with that problem, an autoclave which is the instrument to sterilize the medical
equipments must be paid attention especially in its temperature accurately level. A research
about temperature validity test with time controller in the autoclave that has been measured
30 limes in 4 points which has been determined, also it is done calibration test in the
institution in charge, BPFK (Safety Hall of Medical Facility) Surabaya, the result shows
competent and safety because it is still under the specified standard by Health Department of
Indonesia.
thing to do. All of the equipment related to the patient including to do an operation, before
they're used, it" must be cleaned first. If not, it cans contagious disease to another patient, the
hospital's visitors, even the medicalojjicer itself. In the point, the function of the hospital is a
place to heal the sick people. But hospital can be also become the source of disease because
there are many patient bringing much disease in the hospital, both catching disease and not
catching disease. In spite of that, the sterilization process in a hospital is needed and a must
which have 10 do like the fixed procedure that have been specified.
Related with that problem, an autoclave which is the instrument to sterilize the medical
equipments must be paid attention especially in its temperature accurately level. A research
about temperature validity test with time controller in the autoclave that has been measured
30 limes in 4 points which has been determined, also it is done calibration test in the
institution in charge, BPFK (Safety Hall of Medical Facility) Surabaya, the result shows
competent and safety because it is still under the specified standard by Health Department of
Indonesia.